SEVERITY AND EXPLOITATION CONTEXT

CVE-2018-14781: Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

Review published vectors, EPSS probability and CISA KEV status independently.

Lookup reads saved CVE and CISA snapshots from this site and requests EPSS from FIRST. The CVE source is contacted when a saved record is unavailable or you request a refresh. External providers receive your IP address. Finding notes and saved assessments are not sent by this feature.

Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Published CVE snapshot: 2026-10-22. Record last updated: 2025-05-22T16:33:08.385Z. The live panels refresh separately; this snapshot remains available if a source cannot be reached.

View the CVE Program source record

icscert · CVSS 3.1: 5.3 Medium

Calculated from the published vector. Base score: 5.3. Provider-reported Base score: 5.3.

Vector and metric explanationsCVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
MetricValueMeaning
Attack vector (AV)AThe attacker must share a restricted network context, such as a local segment or an adjacent protocol network.
Attack complexity (AC)HSuccessful exploitation depends on conditions beyond the attacker’s control, requiring additional preparation or execution effort.
Privileges required (PR)NThe attacker needs no privileges before exploiting the vulnerability.
User interaction (UI)NNo other user must participate for exploitation to succeed.
Scope (S)UConsequences stay within resources managed by the same security authority.
Confidentiality (C)NNo confidentiality impact is demonstrated.
Integrity (I)HIntegrity is completely lost, or modification has a direct, serious consequence.
Availability (A)NNo availability impact is demonstrated.

These definitions explain provider selections; they do not independently verify exploitation.

Review vector in calculator

Affected products reported by the provider

  • Medtronic · MMT- 508 - MiniMed pump
  • Medtronic · MMT – 511 pump Paradigm
  • Medtronic · MMT – 512 / MMT – 712 Paradigm x12
  • Medtronic · MMT – 515 / MMT – 715 Paradigm x15
  • Medtronic · MMT – 522 / MMT – 722 Paradigm REAL-TIME
  • Medtronic · MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME
  • Medtronic · MMT – 523 / MMT – 723 Paradigm Revel
  • Medtronic · MMT – 523(K) / MMT – 723(K) Paradigm
  • Medtronic · MMT – 554 / MMT – 754 MiniMed Veo
  • Medtronic · MMT – 551 / MMT – 751 MiniMed 530G

Consult the advisory for affected version ranges, configurations and remediation. A product listing alone does not establish that your installation is vulnerable.

Source references (3)

Interpret severity and exploitation separately

CVSS measures technical severity. EPSS estimates exploitation probability over the next 30 days. CISA KEV lists known exploited vulnerabilities. A missing EPSS result is not zero probability, and absence from KEV does not prove no exploitation. Review your deployment and retain the version, vector and evidence when documenting a finding.