SEVERITY AND EXPLOITATION CONTEXT

CVE-2018-8868: Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

Review published vectors, EPSS probability and CISA KEV status independently.

Lookup reads saved CVE and CISA snapshots from this site and requests EPSS from FIRST. The CVE source is contacted when a saved record is unavailable or you request a refresh. External providers receive your IP address. Finding notes and saved assessments are not sent by this feature.

Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Published CVE snapshot: 2026-10-22. Record last updated: 2025-05-22T18:14:07.710Z. The live panels refresh separately; this snapshot remains available if a source cannot be reached.

View the CVE Program source record

icscert · CVSS 3.1: 6.2 Medium

Calculated from the published vector. Base score: 6.2. Provider-reported Base score: 6.2.

Vector and metric explanationsCVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L
MetricValueMeaning
Attack vector (AV)PRequires physical access to or manipulation of the vulnerable component.
Attack complexity (AC)HSuccessful exploitation depends on conditions beyond the attacker’s control, requiring additional preparation or execution effort.
Privileges required (PR)LThe attacker needs basic user capabilities before exploitation.
User interaction (UI)NNo other user must participate for exploitation to succeed.
Scope (S)CConsequences extend to resources managed by a different security authority.
Confidentiality (C)HAll information is disclosed, or a limited disclosure has a direct, serious confidentiality consequence.
Integrity (I)LSome modification is possible, with limited scope or consequence.
Availability (A)LAvailability is degraded or interrupted to a limited degree.

These definitions explain provider selections; they do not independently verify exploitation.

Review vector in calculator

Affected products reported by the provider

  • Medtronic · 24950 MyCareLink Monitor
  • Medtronic · 24952 MyCareLink Monitor

Consult the advisory for affected version ranges, configurations and remediation. A product listing alone does not establish that your installation is vulnerable.

Source references (2)

Interpret severity and exploitation separately

CVSS measures technical severity. EPSS estimates exploitation probability over the next 30 days. CISA KEV lists known exploited vulnerabilities. A missing EPSS result is not zero probability, and absence from KEV does not prove no exploitation. Review your deployment and retain the version, vector and evidence when documenting a finding.

Browse CVEs from 2018 · All published CVEs · Calculation methodology