SEVERITY AND EXPLOITATION CONTEXT

CVE-2019-10964: Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control

Review published vectors, EPSS probability and CISA KEV status independently.

Lookup reads saved CVE and CISA snapshots from this site and requests EPSS from FIRST. The CVE source is contacted when a saved record is unavailable or you request a refresh. External providers receive your IP address. Finding notes and saved assessments are not sent by this feature.

Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control

Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Published CVE snapshot: 2026-10-22. Record last updated: 2025-05-22T18:29:15.376Z. The live panels refresh separately; this snapshot remains available if a source cannot be reached.

View the CVE Program source record

icscert · CVSS 3.1: 7.1 High

Calculated from the published vector. Base score: 7.1. Provider-reported Base score: 7.1.

Vector and metric explanationsCVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
MetricValueMeaning
Attack vector (AV)AThe attacker must share a restricted network context, such as a local segment or an adjacent protocol network.
Attack complexity (AC)HSuccessful exploitation depends on conditions beyond the attacker’s control, requiring additional preparation or execution effort.
Privileges required (PR)NThe attacker needs no privileges before exploiting the vulnerability.
User interaction (UI)NNo other user must participate for exploitation to succeed.
Scope (S)UConsequences stay within resources managed by the same security authority.
Confidentiality (C)LSome information is disclosed, with limited scope or seriousness of the direct consequence.
Integrity (I)HIntegrity is completely lost, or modification has a direct, serious consequence.
Availability (A)HAvailability is completely lost, or the disruption has a direct, serious consequence.

These definitions explain provider selections; they do not independently verify exploitation.

Review vector in calculator

Affected products reported by the provider

  • Medtronic · MiniMed 508 pump
  • Medtronic · MiniMed Paradigm 511 pump
  • Medtronic · MiniMed Paradigm 512/712 pumps
  • Medtronic · MiniMed Paradigm 712E pump
  • Medtronic · MiniMed Paradigm 515/715 pumps
  • Medtronic · MiniMed Paradigm 522/722 pumps
  • Medtronic · MiniMed Paradigm 522K/722K pumps
  • Medtronic · MiniMed Paradigm 523/723 pumps
  • Medtronic · MiniMed Paradigm 523K/723K pumps
  • Medtronic · MiniMed Paradigm Veo 554/754 pumps
  • Medtronic · MiniMed Paradigm Veo 554CM/754CM pumps

Consult the advisory for affected version ranges, configurations and remediation. A product listing alone does not establish that your installation is vulnerable.

Source references (3)

Interpret severity and exploitation separately

CVSS measures technical severity. EPSS estimates exploitation probability over the next 30 days. CISA KEV lists known exploited vulnerabilities. A missing EPSS result is not zero probability, and absence from KEV does not prove no exploitation. Review your deployment and retain the version, vector and evidence when documenting a finding.