SEVERITY AND EXPLOITATION CONTEXT

CVE-2019-6538: Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

Review published vectors, EPSS probability and CISA KEV status independently.

Lookup reads saved CVE and CISA snapshots from this site and requests EPSS from FIRST. The CVE source is contacted when a saved record is unavailable or you request a refresh. External providers receive your IP address. Finding notes and saved assessments are not sent by this feature.

Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Published CVE snapshot: 2026-10-22. Record last updated: 2025-05-22T19:26:15.126Z. The live panels refresh separately; this snapshot remains available if a source cannot be reached.

View the CVE Program source record

icscert · CVSS 3.1: 9.3 Critical

Calculated from the published vector. Base score: 9.3. Provider-reported Base score: 9.3.

Vector and metric explanationsCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
MetricValueMeaning
Attack vector (AV)AThe attacker must share a restricted network context, such as a local segment or an adjacent protocol network.
Attack complexity (AC)LExploitation does not depend on special conditions beyond the attacker’s control.
Privileges required (PR)NThe attacker needs no privileges before exploiting the vulnerability.
User interaction (UI)NNo other user must participate for exploitation to succeed.
Scope (S)CConsequences extend to resources managed by a different security authority.
Confidentiality (C)NNo confidentiality impact is demonstrated.
Integrity (I)HIntegrity is completely lost, or modification has a direct, serious consequence.
Availability (A)HAvailability is completely lost, or the disruption has a direct, serious consequence.

These definitions explain provider selections; they do not independently verify exploitation.

Review vector in calculator

Affected products reported by the provider

  • Medtronic · Conexus Radio Frequency Telemetry Protocol
  • Medtronic · MyCareLink Monitor
  • Medtronic · CareLink Monitor
  • Medtronic · CareLink 2090 Programmer
  • Medtronic · Amplia CRT-D
  • Medtronic · Claria CRT-D
  • Medtronic · Compia CRT-D
  • Medtronic · Concerto CRT-D
  • Medtronic · Concerto II CRT-D
  • Medtronic · Consulta CRT-D
  • Medtronic · Evera ICD
  • Medtronic · Maximo II CRT-D
  • Medtronic · Maximo II ICD
  • Medtronic · Mirro ICD
  • Medtronic · Nayamed ND ICD
  • Medtronic · Primo ICD
  • Medtronic · Protecta ICD, Protecta CRT-D
  • Medtronic · Secura ICD
  • Medtronic · Virtuoso ICD
  • Medtronic · Virtuoso II ICD
  • Medtronic · Visia AF ICD
  • Medtronic · Viva CRT-D
  • Medtronic · Brava CRT-D
  • Medtronic · Mirro MRI ICD

Consult the advisory for affected version ranges, configurations and remediation. A product listing alone does not establish that your installation is vulnerable.

Source references (2)

Interpret severity and exploitation separately

CVSS measures technical severity. EPSS estimates exploitation probability over the next 30 days. CISA KEV lists known exploited vulnerabilities. A missing EPSS result is not zero probability, and absence from KEV does not prove no exploitation. Review your deployment and retain the version, vector and evidence when documenting a finding.

Browse CVEs from 2019 · All published CVEs · Calculation methodology