SEVERITY AND EXPLOITATION CONTEXT

CVE-2022-38100: Contec Health CMS8000

Review published vectors, EPSS probability and CISA KEV status independently.

Lookup reads saved CVE and CISA snapshots from this site and requests EPSS from FIRST. The CVE source is contacted when a saved record is unavailable or you request a refresh. External providers receive your IP address. Finding notes and saved assessments are not sent by this feature.

Contec Health CMS8000

The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.

Published CVE snapshot: 2026-10-22. Record last updated: 2025-04-16T16:10:19.613Z. The live panels refresh separately; this snapshot remains available if a source cannot be reached.

View the CVE Program source record

icscert · CVSS 3.1: 7.5 High

Calculated from the published vector. Base score: 7.5. Provider-reported Base score: 7.5.

Vector and metric explanationsCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
MetricValueMeaning
Attack vector (AV)NReachable across a network boundary, including internet or routed network paths.
Attack complexity (AC)LExploitation does not depend on special conditions beyond the attacker’s control.
Privileges required (PR)NThe attacker needs no privileges before exploiting the vulnerability.
User interaction (UI)NNo other user must participate for exploitation to succeed.
Scope (S)UConsequences stay within resources managed by the same security authority.
Confidentiality (C)NNo confidentiality impact is demonstrated.
Integrity (I)NNo integrity impact is demonstrated.
Availability (A)HAvailability is completely lost, or the disruption has a direct, serious consequence.

These definitions explain provider selections; they do not independently verify exploitation.

Review vector in calculator

Affected products reported by the provider

  • Contec Health · CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor

Consult the advisory for affected version ranges, configurations and remediation. A product listing alone does not establish that your installation is vulnerable.

Source references (1)

Interpret severity and exploitation separately

CVSS measures technical severity. EPSS estimates exploitation probability over the next 30 days. CISA KEV lists known exploited vulnerabilities. A missing EPSS result is not zero probability, and absence from KEV does not prove no exploitation. Review your deployment and retain the version, vector and evidence when documenting a finding.

Browse CVEs from 2022 · All published CVEs · Calculation methodology